doi:10.22028/D291-36295 | Title: | Annular closure device breakage due to recurrent lumbar disc herniation: a case report |
| Author(s): | Burkhardt, Benedikt W. Oertel, Joachim M. |
| Language: | English |
| Title: | Acta Neurochirurgica |
| Volume: | 163 |
| Issue: | 1 |
| Pages: | 269–273 |
| Publisher/Platform: | Springer Nature |
| Year of Publication: | 2020 |
| Publikation type: | Journal Article |
| Abstract: | Detailed surgical management, magnetic resonance imaging (MRI), and computer tomography (CT) images of a broken annular closure device (ACD) have not been reported yet. In this case, a 28-year-old male presented with a new onset of radiculopathy three years after lumbar discectomy and placement of an ACD. The CT-myelography and MRI revealed a recurrent disc herniation (RDH) and dislocation of a broken ACD. ACD removal was performed and confirmed breakage due to RDH with scarring around the RDH and displaced ACD. Implant-associated complications and management should be reported in detail in order to enhance knowledge on device-related complications. |
| DOI of the first publication: | 10.1007/s00701-020-04651-9 |
| URL of the first publication: | https://link.springer.com/article/10.1007/s00701-020-04651-9 |
| Link to this record: | hdl:20.500.11880/32967 http://dx.doi.org/10.22028/D291-36295 |
| ISSN: | 0942-0940 0001-6268 |
| Date of registration: | 31-May-2022 |
| Faculty: | M - Medizinische Fakultät |
| Department: | M - Neurochirurgie |
| Professorship: | M - Prof. Dr. Joachim Oertel |
| Collections: | SciDok - Der Wissenschaftsserver der Universität des Saarlandes |