Please use this identifier to cite or link to this item: doi:10.22028/D291-40105
Title: Biomechanical Performance of a Novel Implant Design in Simulated Extraction Sites and Sinuslift Procedures
Author(s): Klär, Virgilia
Zimmerer, Rüdiger
Schulz, Annika
Lethaus, Bernd
Language: English
Title: Applied Sciences
Volume: 13
Issue: 13
Publisher/Platform: MDPI
Year of Publication: 2023
Free key words: insertion torque
primary implant stability
extraction socket
sinus lift
implant design
DDC notations: 610 Medicine and health
Publikation type: Journal Article
Abstract: With increasing experience and in an attempt to shorten overall treatment times, implant placement in combination with tooth extractions and sinus lift procedures has become popular. In both cases, primary stability has to be achieved by either engaging apical and oral regions of trabecular bone or by engaging residual host bone beneath the sinus cavity. Extraction sites were formed by pressing a root analog into homogeneous low density polyurethane foam which was used as bone surrogate while a 3 mm thick sheet of medium density foam was used for mimicking a sinus lift situation. Two types (n = 10) of bone level implants with a conventional tapered design and a cervical back taper (NobelActive; control) and a novel design characterized by a shift in core diameter and thread geometry (AlfaGate; test) were placed in these models following conventional osteotomy preparation. Insertion torque was measured using a surgical motor and primary stability was determined by resonance frequency analysis. Statistical analysis was based on Welch two sample t tests with the level of significance set at α = 0.05. In sinuslifting, NobelActive implants required significantly higher insertion torques as compared to AlfaGate (p = 0.000) but did not achieve greater implant stability (p = 0.076). In extraction sites, AlfaGate implants showed both, significantly higher insertion torques (p = 0.004) and significantly greater implant stability (p = 0.000). The novel implant design allowed for greater primary stability when being placed in simulated extraction sockets and sinuslift situations. While in extraction sockets the position of condensing threads in combination with an increase in core diameter is beneficial, the deep cervical threads of the novel implant lead to superior performance in sinuslift situations.
DOI of the first publication: 10.3390/app13137541
URL of the first publication: https://doi.org/10.3390/app13137541
Link to this record: urn:nbn:de:bsz:291--ds-401055
hdl:20.500.11880/36114
http://dx.doi.org/10.22028/D291-40105
ISSN: 2076-3417
Date of registration: 17-Jul-2023
Faculty: M - Medizinische Fakultät
Department: M - Zahn-, Mund- und Kieferheilkunde
Professorship: M - Keiner Professur zugeordnet
Collections:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Files for this record:
File Description SizeFormat 
applsci-13-07541-v3.pdf1,5 MBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons