Please use this identifier to cite or link to this item:
doi:10.22028/D291-46748 | Title: | Assessment of scintigraphic imaging with Terbium-161: A visual and quantitative analysis |
| Author(s): | Rosar, Florian Petto, Sven Ganz, Sebastian Burgard, Caroline Maus, Stephan Bartholomä, Mark Miederer, Isabelle Schreckenberger, Mathias Ezziddin, Samer Schaefer-Schuler, Andrea |
| Language: | English |
| Title: | Applied Radiation and Isotopes |
| Volume: | 226 |
| Publisher/Platform: | Elsevier |
| Year of Publication: | 2025 |
| Free key words: | 161Tb Tb-161 SPECT Quantification Phantom measurement |
| DDC notations: | 610 Medicine and health |
| Publikation type: | Journal Article |
| Abstract: | The Auger emitter 161Tb is an increasingly discussed radionuclide for targeted radionuclide therapy. The aim of this study was to assess the feasibility of scintigraphic imaging with 161Tb in terms of image quality and quantitative capabilities by phantom measurements and to evaluate the suitability of this radionuclide for clinical use. Phantom measurements were conducted using a standardized NEMA IEC body phantom filled with activity concentrations ranging from 2.5 GBq to 100 MBq. Both visual and quantitative analyses were performed, including assessment of the image calibration factor (CF), as well as the recovery coefficient (RC) and the contrast-to-noise ratio (CNR) of the individual spheres. The results observed in this study demonstrate that quantitative SPECT/CT imaging with 161Tb is feasible over a wide range of activity making this radionuclide suitable for clinical applications. Acquiring a total of at least 5 million photopeak counts enables visual detectability of lesions of diameter lower than 20 mm and quantitative calibration for dosimetry purposes. |
| DOI of the first publication: | 10.1016/j.apradiso.2025.112171 |
| URL of the first publication: | https://doi.org/10.1016/j.apradiso.2025.112171 |
| Link to this record: | urn:nbn:de:bsz:291--ds-467486 hdl:20.500.11880/40971 http://dx.doi.org/10.22028/D291-46748 |
| ISSN: | 0969-8043 |
| Date of registration: | 16-Jan-2026 |
| Description of the related object: | Supplementary data |
| Related object: | https://ars.els-cdn.com/content/image/1-s2.0-S0969804325005160-mmc1.docx |
| Faculty: | M - Medizinische Fakultät |
| Department: | M - Radiologie |
| Professorship: | M - Prof. Dr. Samer Ezziddin |
| Collections: | SciDok - Der Wissenschaftsserver der Universität des Saarlandes |
Files for this record:
| File | Description | Size | Format | |
|---|---|---|---|---|
| 1-s2.0-S0969804325005160-main.pdf | 2,83 MB | Adobe PDF | View/Open |
This item is licensed under a Creative Commons License